| MDF-U700VXC-PA |
| Cooling Performance |
| Temperature Setpoint Range |
-50°C to -86°C |
| Dual Compressor Cooling System |
-86°C (ambient temperature -30°C) |
| Single System Cooling Performance (Indefinite) |
> -65°C (ambient temperature -30°C)
|
| |
| Cabinet Design and Construction |
| Cabinet Insulation |
Patented VIP ® PLUS composite wall, energy savings |
Exterior Dimensions
(w x f-b x h) |
39.8" x 34.3" x 79.1"
(1010 x 870 x 2010 mm) |
Interior Dimensions
(w x f-b x h) |
34.3" x 23.6" x 55.1"
(870 x 600 x 1400 mm) |
| Volume |
25.7 cu.ft. (728 L) |
| Shelves |
3, adjustable, stainless steel (4 compartments) |
| Access Port |
Diameter= 0.5" (17 mm); 3 locations |
| Vacuum Release Port |
Standard manual |
| |
| Refrigeration System |
| Refrigerant Technique |
-86° Dual Independent Compressor Cooling Systems with Automatic Switchover |
| Compressors |
2 Compressors, Hermetically 1100W Cool Safe compressor, air-cooled |
| Refrigerant |
Environmentally Safe HFC |
| Air Exchange |
Dual condenser fans with particulate trap; no filters are required |
| Cycle Options |
Standard or energy-saving EcoMode operation |
| |
| Controls, Alarm, Monitoring |
| Controller |
Microprocessor-based, door-mounted control with graphical LCD information center |
| Data Management |
Internal Data log with log file memory |
| Communications |
Data can be transmitted via RS485 interface |
| Alarms |
High/Low temperature, Door ajar, Power failure, Remote Alarm Contact, Part Replacement Notification, Fan Motor Lock, Indivdual Cooling Circuit Check |
| Remote Alarm Contacts |
Standard; Normally Open and Normally Closed; Dry Contacts |
| Electrical |
208V-230V, AC, 60Hz, 1 phase, NEMA 6-15, 15 amp separate breaker recommended
|
| |
| Optional Accessories |
| Digital Temperature Recorder |
Auxiliary Data Logger, HamsterDT2 |
| Liquid CO2 Back-Up System |
Auxiliary tank back in event of power failure, CVK-UB2(I)-PW |
| Liquid N2 Back-Up System |
Auxiliary tank back in event of power failure, CVK-UBN2-PW |
| Chart Recorder |
Circular Chart Temperature Recorder, 7 Day, MTR-C954 |
| Chart Paper |
6" Diameter, 7 Day Chart , C7100386REV |
| Inner Door Configurations |
See icon and key to catalog number |
| Replacement Pen, Red |
Felt tip pens, 6 per pack, R252 |
| Replacement Pen, Blue |
Felt tip pens, 6 per pack, R253 |
| 4-20 MA Output |
Independent Analogue Output MTR-420MAC |
| Touch-Screen Temperature Recorder |
Full Color Touch-Screen Digital Display FT-560SN |